Description
Accuracy:
- Infusion: linear rate ± 3%. Bolus: linear displacement ± 3%.
Battery duration:
- Typical use: 150 hours. Low battery and typical use: 4 hours
Batteries:
- Four Alkaline cells of size “C”.
Operating voltage:
- 5 – 7 Volts.
Operating current:
- 20 mA.
Low battery voltage:
- 5.0 volts approximately.
Flow rates:
- 0 to 600 ml / h.
Flow profile:
- Bolus: continuous.
- Infusion: pulsed intermittently.
Occlusion Force:
- 8 ± 1 lbs.
Maximum occlusion pressure:
- Syringe of 11 psi with 60 cc.
- 20 psi with 20 cc syringe.
Occlusion Detection Time:
- 120 seconds at 36 ml / h. (Reverse related to the flow)
Volume stored in the occlusion:
- 1.1 ml approximately.
Effect on accuracy:
- None at Occlusion Pressure.
Size:
- 9.2 x 4.5 x 2.0 inches.
Weight:
- 2 pounds. With batteries.
Syringes:
- Monoject® or BD® 60 cc.
- Monoject® or BD® 20 cc.