Drager Narkomed MRI

A single vaporizer for the style 19.1 vaporizers and the volume, pressure and oxygen monitor of CORm in first generation units. The second generation units have a Drager GS type screen with volume, pressure and oxygen control. The absorber assembly has the known sensation of Narkomed but is a compatible version of MRI. The units are small and have been tested near the strongest magnets.

  • Fan: AV2 dual circuit pneumatic rising fan + style fan without tidal volume compensation. Volume control modes (VCV), Manual / spontaneous. Mechanical bag / ventilation switch.
  • Machine checklist: FDA manual checklist.
  • Flow meters: Traditional needle valve with glass flow screen.
  • Breathing circuit: Traditional absorbent head. The breathing circuit is of greater volume (two absorbent containers). You can use loose granules or pre-loaded containers. It can be used with nonrespiratory circuits.
  • Electrical power failure: Most functions, except for patient monitoring, are conserved during a power failure, since it is a basic pneumatic-mechanical machine.
Formulario de Contacto Productos
Primer y segundo nombre
Apellidos completos
PARA MAS INFORMACIÓN CONTÁCTENOS AL:

Clic Aquí-Enviar mensaje a WhatsApp

+57 304 355 13 63

+57 4 322 36 03

+1 707 294 8221

Description

general

  • Dimensions:  (W x H x D) 31.5 “x 53” x 27 “(80 x 135.3 x 68.6 cm)
  • Weight:  230 lbs (104.3 kg)
  • Battery charger weight:  3 pounds (1.36 kg)

Electricity

  • 120 Volt Power supply 100 – 120 VAC @ 50/60 Hz
  • 220/240 Volt Power supply 200 – 240 VAC @ 50/50 Hz

Drums

  • Reserve Power Time “3 hours
  • Charging time “16 hours

Gas delivery system

  • Number of gases 3 Gas capacity (O2, N2O, Air)
  • Oxygen discharge rate 55 L / min
  • Low oxygen supply
  • Pressure alarm 34 – 40 psi

Vaporizer

  • Vapor 2000
  • Positions 2
  • Fixed assembly

Fan

  • AV 2+
  • Frequency 1 – 99 (in increments of 1 BPM)
  • I: E Ratio 4: 1 – 1: 4,5
  • Inspiratory Flow 10 -100 L / min
  • Volume Tidal 20 -1500 mL
  • Adjustment of the pressure limit 15 -120 cmH2O

Oxygen monitoring

  • Display range 10 -100%
  • Resolution 1 vol% O2

Monitoring of airway pressure

  • Display range -10 to 125 cmH2O
  • Resolution 1 cmH2O

Monitoring of Tidal Volume

  • Display range .01 – 2.0 L
  • Resolution 0.01 mL

Utility requirements

  • The system works with a 120 VAC source at 50/60 Hz.
  • Regulatory sets and cylinder heads for O2 and N2O external cylinders.
  • O2, N2O and air hoses to connect the piping sources to the connections of the DISS machine.

Tests with MRI scanners up to 3.0 Tesla (shielded or unshielded)

  • Tests performed with the unit touching the magnetic resonance.