Description
- ECG Selection of derivation: STD, pallets, I, II, III
- Entry: Remote ECG through Quick-Look® defibrillator paddles, Fast-Patch® disposable defibrillation / ECG electrodes or three-lead patient cable
- Patient cable length: 13 feet in total length (10 feet with three feet shunt)
- Common mode rejection: 100 dB min. With respect to the ground chassis and minimum of 65 dB. With respect to isolated ground, when measured at 60Hz (the common mode range for the patient’s cable entry is 10 Volts peak with respect to isolated ground)
- Screen size of the monitor: 4 “(W) x 3” (H)
- Sweeping speed: 25 mm / sec
- Frequency response: no diagnostic
- The ECG leads: 1.0 to 40 Hz (-3dB)
- The pallets: 2.2 to 20 Hz (-3dB)
- ECG size: Adjusts the amplitude of the ECG trace on the monitor and the band registration card
- Heart rate meter: Three digital readings on the heart rate screen of 20 to 300 beats per minute, heart rate outside this range do not produce valid systole tones or display of heart rate
- Heart rate alarm: the alarm limits are selectable by the user with three high / low settings (150/40, 120/60, 160/90 – other limit options are available through the configuration menu); 1 mV Cal: the momentary impulse button on the rear panel simulates a signal of 1 mV impulse to the ECG input